방사선치료계획 정도관리 방법에 따른 신뢰도 평가 Reliability estimation about quality assurance method of radiotherapy planning

김정호, 김가중, 유세종, 김기진
  • 언어KOR
  • URLhttp://db.koreascholar.com/Article/Detail/293005
대한안전경영과학회지
제17권 제1호 (2015.03)
pp.119-124
대한안전경영과학회 (Korea Safety Management & Science)
초록

According as radiation therapy technique develops, standardization of radiation therapy has been complicated by the plan QA(Quality Assurance). However, plan QA tools are two type, OADT (opposite accumulation dose tool) and 3DADT (3 dimensional accumulation dose tool). OADT is not applied to evaluation of beam path. Therefore tolerance error of beam path will establish measurement value at OADT. Plan is six beam path, five irradiation field at each beam path. And beam path error is 0 degree, 0.2 degree, 0.4 degree, 0.6 degree, 0.6 degree, 0.8 degree. Plan QA accomplishes at OADT, 3DADT. The more path error increases, the more plan QA error increases. Tolerance error of OADT path is 0.357 using tolerance error of conventional plan QA. Henceforth plan QA using OADT will include beam path error. In addition, It will increase reliability through precise and various plan technique.

저자
  • 김정호(건양대학교병원 방사선종양학과) | Jeong-Ho Kim
  • 김가중(극동대학교 방사선학과) | Gha-Jung Kim Corresponding Author
  • 유세종(건양대학교병원 영상의학과) | Se-Jong Yoo
  • 김기진(건양대학교병원 영상의학과) | Ki-Jin Kim