The evaluation of homogeneity and stability of proficiency testing materials in indoor air quality test
In this study, the homogeneity and stability of standard samples for proficiency testing in indoor air quality within the country (formaldehyde, benzene, toluene, ethylbenzene, p-xylene, styrene, TVOC) were evaluated. The procedures and statistical analysis methods applied in ISO/IEC 13528 (2009) and KS A ISO Guide 35 (2005) were applied as evaluation methods. The homogeneity evaluation was a statistical analysis of repeated measurements of each of the 11 ports and between the 11 ports concentration data. As a result, the coefficient of variation (CV) was within the range of 1.9%~5.9%. The difference between the ports was found to be insignificant and met the statistical standard specified in KS Q ISO 13528. The stability evaluation was assessed by the change in concentration over the long-term stability of the standard samples stored for 90 days. The coefficient of variation (CV), which was within the range of 2.6%~9.0%, exhibited changes in the concentration of the long-term stored standard samples. However, the results satisfy the statistical standard specified in KS A ISO Guide 35. Overall, there is no significant difference between the homogeneity of the standard samples by the port and the stability of the long-term stored samples. Therefore, it is considered to be an appropriate method to supply standard samples in an indoor air quality proficiency test.