논문 상세보기

Monitoring on Dose Index Analyzed in the Mammography KCI 등재

유방촬영검사에서 선량지표분석에 대한 모니터링

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/348833
서비스가 종료되어 열람이 제한될 수 있습니다.
Journal of the Korean Society of Radiology (한국방사선학회논문지)
한국방사선학회 (The Korea Society of Radiology)
초록

유방촬영검사(Mammography)에서 선량지표를 분석하여 한국식품의약품안전처(Korea Food & Drug Administration)에서 제시하는 가이드라인과 비교분석하여 의료방사선의 관리가 잘 이루어지고 있는지 파악하고자 하였다. 방법으로는 유방촬영검사를 시행한 후 의료영상저장정보시스템(Picture Archiving Communication System, PACS)으로 전송되어진 선량 보고서(dose report)와 의료전자차트(Electronic Medical Record, EMR)를 참고하여 양쪽 유방 상하방향촬영(cranio-caudal, CC), 내외 사방향 촬영(medio-lateral oblique, MLO)검사에서 관전압(kVp), 조사시간(exposure time), 관전류량(mAs), 유방압박두께(compressed breast thickness), 평균유선선량(average glandular dose), 체질량지수(Body Mass Index, BMI)를 분류하여 분석하였다. 결과적으로 연령별에 따른 검사부위 CC, MLO에서 유방압박두께는 50-59세에서 45.6 mm, 49.6 mm로 가장 두껍게 나타났으며 전체적인 평균유방압박두께는 CC에서 44.2 mm, MLO 48.9 mm로 MLO가 4.7 mm 더 두껍게 나타났다. 평균유선선량은 CC에서 1.05 mGy, MLO에서 1.14 mGy로 CC보다 MLO에서 0.09 mGy가 더 높게 나타났으며 유방압박두께가 10 mm 증가할수록 CC와 MLO에서는 0.15 mGy, 0.17 mGy가 증가하는 것으로 나타났다. 한국식품의약품안전처에서 제시하는 유방 CC에서 1매 촬영 시 환자가 받는 평균유선선량 1.16 mGy 평가와 비교했을 시 1.05 mGy로 나타났으나 60 mm이상에서는 1.30 mGy로 초과하는 것으로 나타났다. 또한 유방압박두께가 높을수록 체질량지수가 높게 나타났으며 체질량지수에 따른 비만기준 25이상인 경우는 유방압박두께가 50 mm 이상일 경우에 비만인 것으로 나타났다.

This study is tried to determine whether the management of medical radiation is well handled by comparison the guidelines of KFDA(korea food & drug administration) with analysis of dose indicator in mammography. As a method, it is analysed that kVp, exposure time, mAs, compressed breast thickness, average glandular dose and body mass index that were classified in the examination of both breasts by CC(cranio-caudal) and MLO(medio-lateral oblique) with EMR(electronic medical record) and dose report that were sent to the PACS(picture archiving communication system). As a result, in the site inspection according to the age, Compressed breast thickness in CC and MLO were the thickest of 45.6 mm and 49.6 mm in the 50-59 year old respectively. In the overall average compressed breast thickness, CC were 44.2 mm and MLO were 48.9 mm. MLO has more thick by 4.7 mm. In average glandular dose, CC were 1.05 mGy and MLO were 1.14 mGy. MLO has higher by 0.09 mGy than CC. As the compressed breast thickness increases 10mm, CC and MLO increases 0.15 mGy and 0.17 mGy respectively. When it was compared with the average glandular dose of 1.16 mGy per 1 film presented by KFDA, CC was showed 1.05 mGy. However, the 60 mm or more was found to exceed a 1.30 mGy. Also, As the compressed breast thickness was higher, body mass index showed high score. And in the case of 25 or more in the obese body index according to body mass index, it was showed obesity in case of the compressed breast thickness was more than 50mm.

목차
ABSTRACT
 Ⅰ. INTRODUCTION
 Ⅱ. MATERIAL AND METHODS
  1. 연구 대상
  2. 연구 방법
 Ⅲ. RESULT
  1. 연령별에 따른 검사부위별 유방압박두께
  2. 상하방향촬영검사에서 유방압박두께에 따른 매개변수
  3. 내외 사방향 촬영검사에서 유방압박두께에 따른 매개변수
  4. 내외 사방향 촬영검사와 상하방향 촬영검사와의 매개변수의 차이
  5. 유방압박두께에 따른 검사부위별 체질량지수
 Ⅳ. DISCUSSION
 Ⅴ. CONCLUSION
 Reference
 요 약
저자
  • Ji-Hwan Cho(Department of Health Care Clinic, Inje University, Busan Paik Hospital) | 조지환
  • Hyo-Yeong Lee(Department of Radiological Science, Dongeui University) | 이효영
  • In-Chul Im(Department of Radiological Science, Dongeui University) | 임인철 Corresponding Author