A Study on the Change of Image Quality According to the Change of Tube Voltage in Computed Tomography Pediatric Chest Examination KCI 등재

전산화단층촬영 소아 흉부검사에서 관전압의 변화에 따른 화질변화에 관한 연구

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/383237
서비스가 종료되어 열람이 제한될 수 있습니다.
Journal of the Korean Society of Radiology (한국방사선학회논문지)
한국방사선학회 (The Korea Society of Radiology)
초록

흉부 CT 검사의 스캔 기법의 하나인 VOLUME AXIAL MODE를 이용하여 관전압의 변화에 따라 진단적 으로 가치가 높은 영상을 얻기 위하여 화질을 평가하고 적절한 관전압을 제시하고자 한다. CT 장비는 GE 사의 Revolution(GE Healthcare, Wisconsin USA)모델을 이용하였으며, Phantom은 Pediatric Whole Body Phant om PBU-70을 사용하였다. Heart의 SNR 평균차이분석에서는 70 kvp에서 -4.53 ± 0.26 이었고 80 kvp는 -3.34 ± 0.18 이었으며 100 kvp는 -1.87 ± 0.15이었고 70 kvp가 100kvp 보다 약 -2.66정도 SNR이 높았으며 통계적 으로 유의하였다.(p<0.05) Lung의 SNR 평균차이분석에서는 70 kvp에서 -78.20 ± 4.16이었고 80 kvp는 -79.10 ± 4.39이었으며 100 kvp는 -77.43 ± 4.72이었고 70 kvp가 100 kvp 보다 약 -0.77정도 SNR이 높았으며 통계적 으로 유의하였다.(p<0.05). Lung의 CNR 평균차이분석에서는 70 kvp에서 73.67 ± 3.95이었고 80 kvp는 75.76 ± 4.25이었으며 100 kvp는 75.57 ± 4.62이었고 80 kvp가 70 kvp 보다 약 20.9정도 CNR이 높았으며 통계적으 로 유의하였다.(p<0.05) 관전압 100 kvp에서는 70 kvp와 80k vp를 비교 했을 때 심장 영상의 질을 유지하면 서 SNR이 1에 가까웠다. 하지만 70 kvp와 80kvp에서 는 SNR 차이가 없어 70 kvp 로 소아 흉부 CT 검사를 하여 방사선량을 줄일 수 있을 것이다. 반면에 CNR은 70 kvp에서 가장 1에 근사치를 나타냈었으며 80kvp 와 100kvp에서는 차이가 없어 80 kvp로 소아 흉부 CT 검사를 하여 방사선량을 줄일 수 있을 것이다. 또한 Volume Axial mode 검사 시 0.3초의 짧은 scan time으로 검사 할 수 있어서 움직이거나 진정이 필요한 소아 환자에게 유용할 것으로 사료된다.

In short a binary value according to a change in the tube voltage by using one of VOLUME AXIAL MODE of scanning techniques of chest CT image quality evaluation in order to obtain high image and to present the appropriate tube voltage. CT instruments were GE Revolution (GE Healthcare, Wisconsin USA) model and Phantom used Pediatric Whole Body Phantom PBU-70. The test method was examined in Volume Axial mode using the pediatric protocol used in the Y university hospital of mass-produced material. The tube voltage was set to 70kvp, 80kvp, 100kvp, and mAs was set to smart mA-ODM. The mean SNR difference of the heart was -4.53 ± 0.26 at 70 kvp, -3.34 ± 0.18 at 80 kvp, -1.87 ± 0.15 at 100 kvp, and SNR at 70 kvp was about -2.66 higher than 100 kvp and statistically significant (p <0.05) In the Lung SNR mean difference analysis, -78.20 ± 4.16 at 70 kvp, -79.10 ± 4.39 at 80 kvp, -77.43 ± 4.72 at 100 kvp, and SNR at 70 kvp at about -0.77 higher than 100 kvp were statistically significant. (p <0.05). Lung CNR mean difference was 73.67 ± 3.95 at 70 kvp, 75.76 ± 4.25 at 80 kvp, 75.57 ± 4.62 at 100 kvp and 20.9 CNR at 80 kvp higher than 70 kvp and statistically significant (p <0.05) At 100 kvp of tube voltage, the SNR was close to 1 while maintaining the quality of the heart image when 70 kvp and 80 kvp were compared. However, there is no difference in SNR between 70 kvp and 80 kvp, and 70 kvp can be used to reduce the radiation dose. On the other and, CNR showed an approximate value of 1 at 70 kvp. There is no difference between 80 kvp and 100 kvp. Therefore, 80 kvp can reduce the radiation dose by pediatric chest CT. In addition, it is possible to perform a scan with a short scan time of 0.3 seconds in the volume axial mode test, which is useful for pediatric patients who need to move or relax.

목차
ABSTRACT
Ⅰ. INTRODUCTION
Ⅱ. MATERIAL AND METHODS
    1. 사용 장비
    2. 검사 방법
    3. 선량 평가
    4. 영상 평가(정량적 평가)
    5. 통계 분석
Ⅲ. RESULT
    1. 영상평가 결과
Ⅳ. DISCUSSION
Ⅴ. CONCLUSION
Reference
요 약
저자
  • Gu Kim(Department of Radiology, Pusan National University hospital of Yang San) | 김구
  • Gyeong Rip Kim(Gamma knife center, Pusan National University Hospital of Yang San) | 김경립
  • Soon Ki Sung(Gamma knife center, Pusan National University Hospital of Yang San) | 성순기
  • Jong Hyeok Kwak(Department of Radiology, Pusan National University hospital of Yang San) | 곽종혁 Corresponding Author