n this research, the dose rate was measured using backpack-type scan survey device at 4 sites on Jeju Island, and the radioactivity ratio for each nuclide was evaluated using an high-purity germanium (HPGe) detector. As a result of measuring the dose rate with a backpack-type scan survey device, the average dose rate was the lowest in the measurement site 3 at 0.049 Sv/h, and the highest in the measurement site 1 at 0.066 Sv/h. The average dose rate of the 4 sites on Jeju Island was 0.056 Sv/h, and the dose rate on Jeju Island was lower than that of other regions. The data acquired by scan survey were interpreted using classed post and gridding function of surfer program. The radioactivity ratio of each nuclide in the gamma spectrum measured by the HPGe detector was the highest for K-40 with an average of 87.62%, and the lowest for Pb-214 with an average of 0.63%. In the case of the Jeju Island site, Cs-137 was detected, and the average radioactivity ratio of Cs-137 was 3.27%, which was the background level. The results of this research can be used as basic data on the radioactivity ratio for each nuclide and dose rate at the Jeju Island site. Further studies on the assessment of dose rates and radioactivity ratios in other regions are needed.