We conducted multi-elements determination of reference material certified by the Inorganic Ventures, IV-26, using iCAP 7400 ICP-OES of Thermo Fisher Scientific. And we statistically evaluated analysis results by introducing the in-house proficiency evaluation method implemented at the Ministry of Food and Drug Safety. Ca, Co, Fe, Mg, Ni, and V were selected as target elements, and extended uncertainty was estimated at a confidence level of about 95% and coverage factor k = 2. Five parameters incurred at manufacturing process (standard solution, calibration curve, repeated measurement and dilution factor of the test sample) were considered when determining the uncertainty. En-score can be calculated using the formula En=(x-X)/(Ulab 2+Uref 2)1/2 described in KS Q ISO 13528, where x, Ulab, X, and Uref are the test results, the uncertainty of the result, and the certified value and the uncertainty of the value. And if the absolute value |En| is less than 1, it can be evaluated as a satisfied value. As a result of ICP-OES analysis, each concentration of the elements to be measured was almost similar to the certified concentration of the reference material, and the uncertainty was slightly different. Also since evaluation on multi-elements determination had an En-score within 1, it was confirmed that the analysis results satisfied En evaluation.