This study was performed to evaluate the separation of Sr, Cs, Ba, La, Ce, and Nd using gas pressurized extraction chromatography (GPEC) with anion exchange resin for the quantitation of Neodymium. GPEC is a micro-scaled column chromatography system that provides a constant flow rate by utilizing nitrogen gas. It is overcome the disadvantages of conventional column chromatography by reducing the volume of elution solvent and shortening the analysis time. Here, we compared the conventional column chromatography and the GPEC method. The whole analysis time was decreased by nine times and radioactive wastes were reduced by five times using the GPEC system. Anion exchange resin 1-X4 (200~400 mesh size) was used. The sample was prepared at a 0.8 M nitric acid in methanol solution. The elution solvent was used at a 0.01 M nitric acid in methanol solution. Finally the eluate was analyzed by ICP-MS to determine the identification and recovery. In this case, we applied the natural isotopes of LREEs (139La, 140Ce, and 144Nd) and high activity nuclides (88Sr, 133Cs, and 138Ba) instead of radioactive isotopes for the preliminary test; as a result, unnecessary radioactive waste was not produced. The recoveries were 93.9%, 105.9%, 91.9%, 47.6%, 35.9%, and 79.9% of Sr, Cs, Ba, La, Ce, and Nd, respectively. The reproducibility of recoveries by GPEC were in the range 2.8%–10.9%.