논문 상세보기

혈액 이송을 위한 열전소자 저장용기에 관한 연구 KCI 등재

A Study on Thermoelectric Element Storage Containers for the Transport of Whole Blood Products

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/437774
구독 기관 인증 시 무료 이용이 가능합니다. 4,000원
한국기계기술학회지 (Journal of the Korean Society of Mechanical Technology)
한국기계기술학회 (Korean Society of Mechanical Technology)
초록

In this paper, we developed a thermoelectric storage container that maintains whole blood temperature and studied its characteristics. Using a temperature sensor, we compared the internal temperature with the set temperature in real time and maintained the temperature between 1 and 6°C. In addition, we installed insulation inside the container and used a lithium-ion battery to power the system to ensure a stable whole blood temperature even during long-distance transportation. Instead of a blood pack, a 350-g water bottle was used and the experiment was conducted at an ambient temperature of 25°C. After cooling for 10 minutes, the internal temperature of the container reached 6°C and the battery operated for 60 minutes. In the future, if we improve the thickness and thermal conductivity of the insulation installed in the thermoelectric storage container and utilize phase change materials as a supplement, we will be able to transport more diverse blood products and special reagents stably.

목차
Abstract
1. 서 론
    1.1 연구목 적
    1.2 연구 배경
2. 실험장치
    2.1 실험장치의 원리
    2.2 실험장치의 구성
3. 결과 및 토론
4. 결 론
References
저자
  • 김도훈(Department of Mechanical System Eng., Graduate School, TongMyong Univ. in Pusan.) | Do-Hoon Kim Corresponding author
  • 구학근(Professor, Department of Refrigeration and Air-Conditioning Eng., TongMyong Univ.) | Hak-Keun Ku