검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 2

        1.
        2023.05 구독 인증기관·개인회원 무료
        For the final disposal of radioactive waste, concentration of gamma nuclides such as Co-58, Co-60, Cs-137, Nb-94 have to be determined to meet nuclear regulatory requirements. In general, gamma nuclide analysis can be performed with simple sample pretreatment without complicated chemical separation processes due to the characteristics of the nuclide and high resolution of the measuring equipment. However, when the concentration of Co-60 is high in a specific radioactive waste generated at the NPP, the background is increased by the compton continuum of Co-60. That makes it difficult to evaluate accurately Nb-94, which is in the lower energy band than the gamma ray energy region of Co-60 and especially Cs-137, which is used as a key nuclide of scaling factor. In this study, We consider the problem of MDA dissatisfaction or overestimation due to the increased background by Co-60.
        2.
        2018.04 KCI 등재 서비스 종료(열람 제한)
        본 연구는 점 선원인 60Co, 137Cs 및 혼합부피선원을 이용하여 피크 대 컴프턴 비율, 연속체 배경 스펙트럼을 감소시키기 위해 저 에너지 peak부터 고 에너지 peak에서 측정된 측정치와 PENELOPE와 비교하였다. 또한, 저에너지 부근에서의 변화를 통해 후방산란, 컴프턴 단(compton edge)의 효율 변화를 PENELOPE와 비교하였다. 혼합부피 선원에서 나온 결과를 토양시료에 적용하여 억제와 비 억제(unsuppressed)모드에서 토양시료의 최소검출한계치가 얼마큼 감소하였는지 확인하고자 한다. 60CO(1,173 keV)의 저에너지 영역의 컴프턴 억제가 상당히 되었으며, 137Cs(661 keV) 피크에 대한 Compton edge의 RF는 2.8이다. 특히, 60Co 선원은 1,173.2keV와 1,332.5 keV의 coincidence 감마선을 방출하므로 컴프턴 억제는 대략 21% 감소하였다. 60Co 선원에서 방출되는 1,173keV와 1,332keV의 compton edge의 RF는 3.2, 3.4였으며 피크대 컴프턴 edge 비율은 8:1로 향상되었다. 그리고, PENELOPE와 비교했을 때 불확도는 2% 이내로 잘 일치하였다. Compton unsuppressed 모드에서 661 keV, 1,173 keV 및 1,332 keV의 MDA 값은 각각 0.535, 0.173 및 0.136Bq/kg이었으나, Compton suppressed 모드에서는 0.121, 0.00826 및 0.00728B/kg로 감소하였다. 따라서, Compton suppres sed는 배후방사능과 검출기 자체에 함유된 방사능을 줄일 수 있었다.