Passive sampler is a simple and cost-effective measuring equipment for ambient and indoor air pollution. We studied the performance of a short term (1 hour mean concentration) ozone passive sampler which was coated with a colorant (indigo carmine) to a filter substrate. Acetone and sulfamic acid added ozone passive sampler was investigated to measure short term mean ozone concentration. Ozone response and interference of criteria air pollutant(SO2, NO2, CO) on a short term ozone passive sampler was tested through experimental chamber. The results show sulfamic acid added passive ozone sampler have good response in ozone exposure. Interference of NO2 gas is larger than other two criteria gases.