검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 1

        1.
        2012.11 KCI 등재 서비스 종료(열람 제한)
        The versatile aerosol concentration enrichment system (VACES) have proven useful for providing elevated levels of atmospheric aerosol to human and animal exposures. In this study, we describe a VACES and tests conducted to both optimize the enhancement factor (EF) and characterize how it depends on experiment conditions. Particle number concentrations were measured from upstream and downstream of the system by scanning mobility particle sizer (SMPS) with a long differential mobility analyzer (DMA) in combination with a condensation particle counter (CPC). SMPS was used for to determine VACES particle EF. Particle EF tends to increase for higher the saturator temperature (TSat) and lower the condenser temperature (TCon). TCon higher than 0 °C and TSat lower than 50 °C was the best to obtain the most increase in particle concentration. Correlation analysis of EF with factor variables of TSat and TCon resulted in correlation 0.662 and 0.416, respectively. With all five predictor variables included in a multiple regression model, the EF had a liner correlation with R2 = 0.643.