        Considering the Fukushima nuclear accident and the marine discharge plan of contaminated (or treated) water, it is necessary a seafood monitoring system for radioactive material screening. Currently, radioactivity tests in seafood are conducting in Korea. Although current method using a HPGe detector can provide very low uncertainty in determining radioactivity, there is a limitation in that rapid inspection cannot be performed because of a time-consuming pretreatment process as well as long measurement time (typically 10,000 s). To overcome this limitation, we are developing an insitu inspection device, a kind of screening system, which can monitor the radioactivity in seafood in a near real-time basis. In this study, the actual seafood with a check source was measured to verify the reliability of the Monte Carlo simulation model. The detector used in the experiment was a 5-cm-thick polyvinyl toluene (PVT) plastic scintillator with a 0.5-cm-thick lead shield for background reduction. A Cs-137 check source was placed within seafood. The seafood used in the experiment was fishcake, raw oyster, and dried laver, which is the representative seafood of fish, shellfish, and seaweed. These three seafood products of the same size and shape as the manufacturing process were used to predict the performance realistically. We compared the energy spectrum of the Cs-137 check source obtained from measurements and simulations. The region of interest (ROI) around the Compton edge was set to reduce the influence of the background and electronic noise. The results showed that the measured and simulated spectrum were in good agreement.