초록

Recently, Clinical reports about reducing the delivery time, pain, instrument delivery, and C-section by increasing intrauterine pressure through giving a steady pressure on fundus have been released. However, the use of fundal pressure during the labor is a cause of great concern about medical malpractice and accidents caused by overpressure on fundus. To practice fundal pressure in safe way, medical device is designed. The device consists of inflatable belt, controller, and TOCO transducer; it inflates the belt to give air pressure onto fundus. In this study, we propose the shape of belt to give an air pressure on fundus equally, efficiently, and safely.

목차
Abstract
 1. 서 론
 2. 분만 유도 장치
  2.1 분만유도장치의 구성 및 효과
  2.2 벨트의 형상
 3. 벨트의 유한요소해석
  3.1 벨트의 유체해석
  3.2 공기주머니의 유체-고체연계 해석
  3.3 유체해석과 유체-고체연계 해석 결과 비교
 4. 벨트 압력의 실험
  4.1 센서 보정 및 실험 준비
  4.2 시험 결과 분석
 5. 결 론