The contents of this paper is to develop a passive sampler for H2S measurement. When the H2S gas exists in the air, AgNO3 solution coated filter of white used phenomenon which is exchanged with black. If the H2S gas concentration increased, the color of AgNO3 solution coated filter is discolored more black. H2S passive sampler measures the H2S gas concentration by changed color of AgNO3 solution coated filter. The reproducibility of the H2S passive sampler is very stable to within an error 5%RSD. The black color of AgNO3 solution coated filter showed a linear relationship with the H2S gas concentrations. In addition, correlation of the developed CDM and CR-10(Minolta, Japan) showed a high correlation to 0.99. Manufactured H2S passive sampler must be kept refrigerated, stability and reactivity was observed for up to 20 days