사지의 방사선 치료에 사용하는 water bolus의 적용 시 치료부위의 위치 변화에 따른 방사선 선량분포의 민감도 연구 KCI 등재

A Study on Dose Sensitivity according to Position Variance in Extremities Radiation Therapy with Water Bolus

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/310713
구독 기관 인증 시 무료 이용이 가능합니다. 4,000원
대한안전경영과학회지 (Journal of Korea Safety Management & Science)
대한안전경영과학회 (Korea Safety Management & Science)
초록

The purpose of this study was to investigate radiation dose sensitivity due to displacement of human extremities in the water bolus box on radiation therapy. Water bolus box and human thigh with femur bone were constructed in computerized radiation therapy planning system to verify the absorbed dose. Two 6MV X-ray beams were irradiated bilaterally into water bolus box and then radiation dose were calculated each situation at displacement of middle axis of thigh from the center in water bolus box to right and left direction. Absorbed dose of thigh and femur bone increased by the distance of displacement. The maximum dose of thigh even increased 20% over than prescribed dose. This is in contrast to conventional concept of dose distribution in water bolus box. Based on this result, displacement of body site in the water bolus box have to be averted during radiation therapy.