Korea Atomic Energy Research Institute (KAERI) has been operating the Post Irradiation Examination Facility (PIEF). The facility has many PIE equipment and one of them is a hydrogen analyzer for measuring hydrogen contents in Zr cladding of spent fuel. The cladding tube of fuel is oxidized in the core environment of high temperature and pressure and absorbs some of the hydrogen generated during the oxidation. The hydrogen content increases with the increase of burn-up, and causes hydriding of the material, which degrades the mechanical properties. Therefore, hydrogen content analysis of the cladding tube is required for the performance and integrity evaluation of spent fuel. In PIEF, the hydrogen analyzer extracts hydrogen gas from Zr cladding by the hot extraction method. The hydrogen gas flows with inert gas and oxidizes to H2O through a CuO reagent. Finally, an IR detector measures the hydrogen amount from the absorbed IR intensity at a specific wavelength. Because the equipment is in the glove box and has some consumable parts, the maintenance work was performed as a radiation work.