Improved Algorithm for Gamma- and X-spectroscopic Analysis of Iodine-129 Considering of Energy-Efficiency Correation

  • 언어ENG
  • URLhttps://db.koreascholar.com/Article/Detail/431561
모든 회원에게 무료로 제공됩니다.
한국방사성폐기물학회 학술논문요약집 (Abstracts of Proceedings of the Korean Radioactive Wasts Society)
한국방사성폐기물학회 (Korean Radioactive Waste Society)
초록

I-129 is one of the imporant nuclides that must be determined in the disposal process of radioactive waste in many countries. This radionuclide emits gamma-ray and x-ray photons within the energy range of 29 to 39 keV, consequently, an x-ray detector with high resolution performance is required for the analysis of I-129 activity. An n-type coaxial HPGe detector exhibits higher efficiency characteristics compared to a planar-type HPGe detector, however, its resolution is lower than a planar type. So it is difficult to completely deconvolute and fit the gamma-ray and xray peaks in the spectrum using a general gamma-ray spectrum analysis program such as GammaVision. To address this problem, in a previous study introduced the developed algorithm for the fitting and analysis of I-129 gamma-ray and x-ray spectum by fixing their emission ratios. In this study, we improved the algorithm by considering the variation of the efficiency in the HPGe spectrum, which reflects the actual HPGe crystal condition. And algorithm tests were performed using measured I-129 sample spectra with interfering nuclides acting as background curve are introduced.

저자
  • Junhyuck Kim(Korea Atomic Energy Research Institute (KAERI)) Corresponding author