검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 3

        2.
        2022.05 구독 인증기관·개인회원 무료
        There are three questions arise in radioactivity measurements: (1) Dose the measured value originate from the radioactivity being present in the sample? (2) Is the measurement procedure suitable for the intended measurement purpose with respect to the requirements? (3) What is a range of values fairly sure our true value lies in with a specified probability? These three questions are answered by determining characteristic limits (decision threshold, detection limit and limits of the coverage interval), which are widely used as part of quality assurance in radioactivity measurements. In the past, numerous papers have focused on the questions in different ways, and have drawn a variety of conclusions about the meaning of the different characteristic limits using various terms and symbols. In recent years, substantial efforts were made in order to obtain a systematic and unified way to calculate and express these limits. As a result, the ISO 11929 Series which specify a procedure for calculation of the characteristic limits have been developed. This paper is focused on the calculations of characteristic limits for noble gas monitor (NGM204 monitor) that offers the continuous measurements of radioactive noble gases discharged from the stacks of the HANARO reactor facility. The calculations are based on the standard ISO 11929 as well as the traditional formulas provided by NUREG 1576, ANSI N42-18, ANSI N42-17 and DIN 25482. A comparison is made among the results obtained from the formulas given in each literature.