검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 1

        1.
        2017.12 KCI 등재 구독 인증기관 무료, 개인회원 유료
        The purpose of this study is to set the method condition of the DPRA in Korea throughout reproducibility study. We conducted intra-lab (triplicate) and inter-lab (three labs of CRI, KTR and CU) validation using 20 chemicals (10 chemicals for the proficiency test listed in the OECD test guideline 442C and an additional 10 more chemicals from reference papers). The data from all three labs met the acceptance criteria. Upon intra-lab validation, two positive chemicals out of 20 total chemicals showed false negative, and one negative chemical showed false positive. In inter-lab validation of three labs, the sensitivity data were 83, 83 and 88% each, and the specificity data was 100, 100 and 88% each. So the accuracy of the three labs was equal to 90%. During these studies, we also checked and improved various limitations that arose. Taken together, the results indicated that the DPRA proposed by OECD test guideline 442C is expected to become a well-established method under Korean GLP applied system.
        4,000원