검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 3

        3.
        2023.11 구독 인증기관·개인회원 무료
        NFDC (Nuclear Fuel and materials Data Center) is designated as a one of the data center of National Standard Reference Center from Ministry of Trade Industry and Energy at Dec. 30 2008. The fields of designation were nuclear fuel and energy materials. NFDC produces standard reference data of nuclear fuel and materials. To ensure reliability of experimental data uncertainty should be estimated. There are two kinds of uncertainty: A-type uncertainty from tester and B-type uncertainty from experimental equipments. To reduce the former, the measurement should be repeated for sufficient amount of times, and to reduce the latter type uncertainty all equipment have to be calibrated. In this study self calibration process of thermo-mechanical analyzer (TMA) was established to ensure the B-type uncertainty. The self calibration was performed using the standard reference material and correction factor was obtained. The correction factor was defined as the ratio of the thermal expansion value of the standard reference material reported in the certificate and the thermal expansion value measured using TMA. It is believed that the uncertainty evaluation process of TGA data developed in this study will be helpful for increasing reliability and stability evaluation of nuclear fuel and spent fuel.