검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 1

        1.
        2001.09 KCI 등재 구독 인증기관 무료, 개인회원 유료
        In-shoe measurement systems allow the clinician and researcher to examine the pressure parameters within the shoe. The purpose of this study was to investigate the test-retest reliability of plantar pressures using the Parotec system over speeds and plantar regions. Seventeen healthy subjects were recruited for the study. Sampling rate was 100 Hz, and data of six variables (pressure on medial heel, lateral heel, 1st metatarsal head, 5th metatarsal head, and great toe and total impulse) were collected in four different gait speed (1.0 m/sec, 1.5 m/sec, 2.0 m/sec, and comfortable walking speed) in each day. The result indicates fair to excellent reliability between the two day test. Intraclass correlation coefficients (ICCs) ranged from .693 to .979, and range of reliability was similar depending on the speed and plantar region. In most cases, data recorded by the Parotec systems provide good evidence for the reliability.
        4,000원