        검색결과 1

        1.
        2010.04 구독 인증기관 무료, 개인회원 유료
        The purpose of this study was to investigate the actual conditions of radiation safety supervision in animal clinics using quality assurance (QA) and quality control (QC) of diagnostic X-ray units. The surveys for QA/QC, equipment condition, and safety supervision were carried out in 18 animal clinics randomly. The QA/QC included reproducibility of dose exposure, kVp, mAs, collimator accuracy test, collimator luminance test, X-ray view box luminance test, grounding system equipment test and external leakage current test. As a result, 44.44% of reproducibility of dose exposure was proper, 81. 25% of kVp test was good, and 100% of mAs test was appropriate. Also, 66.66% of collimator accuracy test was proper, 61.11% of collimator luminance test was good, 53.13% of X-ray view box luminance test was suitable. In addition, only 5.55% of grounding system equipment and ground resistance was proper, 63.64% of external leakage current test was appropriate in grounding system equipment test.