편향전자석의 자장변화에 따른 광자선 선질평가에 관한 고찰 KCI 등재

Study of photon beam quality tool at magnetic field change in bending magnet

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/284760
구독 기관 인증 시 무료 이용이 가능합니다. 4,000원
대한안전경영과학회지 (Journal of Korea Safety Management & Science)
대한안전경영과학회 (Korea Safety Management & Science)
초록

Beam quality is changed about magnetic field of bending magnet. Evaluation of beam quality using PDD(Percentage Depth Dose) at 10cm depth at recommendation of AAPM(America Academy of Pain Medicine). However this evaluation shows fragmentary element. Therefore this study is applied to three value, 10cm divided by 5cm depth PDD, 20cm divided by 10cm depth PDD, 30cm divided by 20cm depth PDD, at change the magnetic field. PDD is measured at magnetic field changed ±1%, ±2% at 6MV(Mega Voltage), 10MV photon. The plan technique is 3 portal plan using Core-Plan at human pelvic phantom. Conventional and presented methods are compared at maximum and minimum dose. The presented method increased discernment of relieve the unequal distribution and energy area than conventional method. Henceforth, application of presented method will be considered. Development of energy measurement method and detector miniaturization will be needed about continuous study.

목차
Abstract
 1. 서 론
 2. 대상 및 방법
  2.1 대상
  2.2 방법
 3. 결 과
  3.1 선량평가 오차비교
  3.2 선량평가 기준비교
  3.3 오차에 따른 기준비교
 4. 고찰 및 결론
 5. Reference