임산부 흉부촬영 시 복부차폐의 적정성 평가 KCI 등재

Adequacy Assessment to Abdomen Shield of Pregnant X-ray Chest PA

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/310714
구독 기관 인증 시 무료 이용이 가능합니다. 4,000원
대한안전경영과학회지 (Journal of Korea Safety Management & Science)
대한안전경영과학회 (Korea Safety Management & Science)
초록

When performing Chest x-ray examination to pregnant woman, normally we shield back side of abdomen. In this situation, scattered rays made by equipment and surrounding structure can enter front side of abdomen. Therefore, in this study, we evaluate suitability of abdomen shield especially to pregnant woman. In case of One shielding material placed back of abdomen, the measured value is 0.676±0.19 uSv/hr. Two shielding material is 0.764±0.04 uSv/hr. Three is 0.685±0.16 uSv/hr. The exposure dose inferred in this study does not excess annual effective dose limit. But It is not mean absolute safety. So we have to minimize occurrence of stochastic effect of radiosensitivity by shielding front side of abdomen of pregnant woman in clinic.

저자
  • 김기진(건양대학교병원 영상의학과) | Ki-Jin Kim
  • 김가중(극동대학교 방사선학과) | Gha-Jung Kim Corresponding Author