        1.
        2015.12 KCI 등재 구독 인증기관 무료, 개인회원 유료
        When performing Chest x-ray examination to pregnant woman, normally we shield back side of abdomen. In this situation, scattered rays made by equipment and surrounding structure can enter front side of abdomen. Therefore, in this study, we evaluate suitability of abdomen shield especially to pregnant woman. In case of One shielding material placed back of abdomen, the measured value is 0.676±0.19 uSv/hr. Two shielding material is 0.764±0.04 uSv/hr. Three is 0.685±0.16 uSv/hr. The exposure dose inferred in this study does not excess annual effective dose limit. But It is not mean absolute safety. So we have to minimize occurrence of stochastic effect of radiosensitivity by shielding front side of abdomen of pregnant woman in clinic.
        4,000원
        2.
        2012.12 KCI 등재 서비스 종료(열람 제한)
        다양한 질환이 발생 될 수 있는 임신중에 산모의 건강상태를 검사하기 위해서는 임산부의 X-ray 흉부촬영이 필수적으로 이루어지게 된다. 따라서 임산부의 X-ray 흉부촬영에 관한 의료용 방사선의 사용은 기본적인 검사로 사용되어지고 있다. 따라서 임신초기인 착상기 부터 태아의 각종 장기 형성 및 기관 형성기까지 대부분의 임신기간은 방사선의직접적인 차폐가 요구되어진다. 하지만 현재 사용되고 있는 납치마는 임산부가 직접 손으로 잡고 있고 는 불편함으로인한 산모의 정확한 X-ray 흉부촬영 자세가 이루어지지 않고 있는 실정이다 또한 이러한 결과로 인해 산모의 X-ray흉부촬영 영상이 폐첨부와 견갑골이 자주 겹쳐 흉부영상이 질적으로 저하되는 현상을 가져다 주고 있다. 이러한 문제점을 극복하기 위해 산모용 납치마를 제작하였다. 결과적으로 새로 디자인한 산모용 납치마는 산모의 X-ray흉부 촬영시 편한 자세를 유지 할 수 있게 해주었다. 아울러 산모가 정확한 자세를 취 할 수 있게 되었다. 더 나아가서 산모의 X-ray흉부 촬영시 견갑골과 폐첨부가 겹치는 현상을 최소화 해주어서 산모의 흉부촬영 영상에 최적화된 이미지를 제공 할 수 있게 해 주었다.