For safe disposal of radioactive wastes, accurate analysis of nuclear isotopes is important. It is known that there are 14 nuclides that have to identify nuclide-specific concentration levels. 63Ni, one of non-volatile nuclear isotopes which is included in those 14 nuclides, has to follow chemical separation for exact analysis. As various analysis methods were developed, various methods for analyzing 63Ni also emerged. Past method has used measurement specimens of 59Ni, after 59Ni measurement has been done. It used HClO4, known as strong oxidizing agent, to dissolve DMG, an organic substance used to form 59Ni precipitates. Nowadays, we analyze 59Ni and 63Ni simultaneously, which enables short analysis time, without use of HClO4. But high accuracy is just as important as short measurement time and efficiency. So, this paper compare 63Ni specific activity value used new method with the value, past method used, using real sample’s data. As a result, all sample data from new method’s relative 63Ni specific activity is within the uncertainty range of past ones based on past specific activity value. Consistency of new method’s result and past method’s data increased the reliability of the data and accuracy of those methods.