검색결과

검색조건
좁혀보기
검색필터
결과 내 재검색

간행물

    분야

      발행연도

      -

        검색결과 1

        1.
        2014.11 구독 인증기관 무료, 개인회원 유료
        We performed the quantitative analysis using MATLAB for slice thickness, spatial resolution, and low contrast resolution with the 50 received images which have suitable judgement by scanning AAPM phantom using GE, PHILIPS, SIEMENS, and TOSHIBA. Slice thickness and spatial resolution were measured by using full width at half maximum(below FWHM). Spatial resolution was confirmed that FWHM interval more than 1.0 mm size. Low contrast resolution was measured that how close by using Roundness Index(below RI) until 6.4 mm size. The spatial resolution 9 images and low contrast resolution 10 images which have disapproved judgement were also had the quantitative analysis. The statistic analysis was judged to have statistically significant differences when p-value is less than 0.05 through SPSS statistics 19.0 and T-test (paired t-test). For slice thickness 5 mm, the average qualitative evaluation group shows 4.98 mm and quantitative evaluation group shows 5.02. For slice thickness 10 mm, the average qualitative evaluation group shows 9.98 mm and quantitative evaluation group shows 9.86. No significant differences (p=0.07, p=0.06) with 5 mm and 10 mm. It confirmed the interval for all FWHM until 1.75, 1.50, 1.25, 1.00 mm as approved images on spatial resolution ,and the average distance was 0.102 mm and the minimum distance 0.054 mm. For disapproved images, all interval was not showed under 1.00 mm. For low contrast resolution, spproved image’s RI value shows average 0.81 and minimum 0.77. The disapproved image’s RI average shows 0.70. Statiscally significant differences were presented (all p<0.05) following hole size.
        4,500원