논문 상세보기

HEDS-IC를 이용한 환경대기 중 암모니아 연속측정에 관한 연구 KCI 등재

  • 언어KOR
  • URLhttps://db.koreascholar.com/Article/Detail/239601
구독 기관 인증 시 무료 이용이 가능합니다. 4,000원
한국냄새환경학회지 (Journal of Korean Society of Odor Research and Engineering)
한국냄새환경학회 (Korean Society Of Odor Research And Engineering)
초록

The ammonia in ambient were sampled by high efficiency diffusion scrubber (HEDS) and analyzed by IC. Ammonia showed high linearity (R²>0.999) of the calibration curve and good repeatability (RSD<5%). The detection limit of Ammonia was about 0.05 ppbv. Average concentration of Ammonia was 12.7 ppbv, Instantaneous maximum concentration was 83.4 ppbv. Continuous sampling method is proper to monitor ammonia which is the odor material instantaneously increased mainly affected by meteorological condition. The sampling and analysis process can be automated and performed in real-time by continuous sampling of HEDS-IC system.

목차
ABSTRACT
요약문
1. 서론
2. 연구 방법
3. 결과 및 고찰
4. 결론
참고문헌
저자
  • 이민도
  • 임용재
  • 이상욱
  • 서석준
  • 서민정
  • 성민영
  • 한진석