초록

In this study, we quantitatively analyzed the data by measuring the radiation shielding rate and uniformity in order to evaluate the performance of an Apron. In addition, storage conditions were also evaluated. The uniformity measurement was performed by evaluating the Apron DICOM images using a PACS program. The experiment was intended for 51 Aprons being used in three hospitals in the Daejeon area. The radiation shielding rate and uniformity were measured per lead equivalent for 0.25 mmPb, 0.35 mmPb, and 0.5 mmPb. As a result, the higher lead equivalents were, the greater differences in the non-uniformity between the top part and the bottom part became (p=0.020). In all hospitals, regarding the non-uniformity of four places in Aprons, all showed statistically significant differences (p<0.01). The average value of the transmitted radiation dose showed less difference (p = 0.005) in the bottom right than in the upper right but was statistically significant. There have been no marks of manufacturing date or the date of purchase in the Apron.

저자
  • 유세종(건양대학교병원 영상의학과) | Se-Jong Yoo
  • 임창선(건양대학교 방사선학과) | Chang-seon Lim Corresponding Author
  • 심규란(건양대학교 방사선학과) | Kyu-ran Sim