논문 상세보기

Validation of ISOCS Measurements Using MicroShield

  • 언어ENG
  • URLhttps://db.koreascholar.com/Article/Detail/431365
모든 회원에게 무료로 제공됩니다.
한국방사성폐기물학회 학술논문요약집 (Abstracts of Proceedings of the Korean Radioactive Wasts Society)
한국방사성폐기물학회 (Korean Radioactive Waste Society)
초록

In nuclear power plant environments, the analysis of gamma-emitting waste materials with complex shapes can be challenging. ISOCS (In-Situ Objective Counting System) is employed to measure the gamma-emitting radionuclide concentrations. However, it is crucial to validate the accuracy of ISOCS measurements. This study aims to validate the accuracy of ISOCS measurement results for spent filters. The ISOCS measurement process begins with modeling and efficiency calculations of the target spent filters using ISOCS software. ISOCS offers the advantage of direct measurement assessment by incorporating shielding materials and collimators into the detector efficiency calculation during the modeling process, without the need for separate efficiency correction sources. To validate the accuracy of ISOCS measurement results, the measured radioactivity values were used as input data for the MicroShield computer code to derive dose rates. These dose rates were then compared to the dose rates measured on-site, confirming the reliability of ISOCS measurements. In the field, ISOCS gamma measurements and surface dose rates were measured for three Cavity filters and four RCP Seal Injection filters. The measured dose rate for the Cavity filters was around 270 Svhr, and the computed values using MicroShield showed an error of approximately 12%. Despite modeling and calculation errors in computer analysis and potential uncertainties in the measurement environment and instrument, the computed values closely matched the measured values. However, the measured dose rate for the RCP Seal Injection filters ranged 2.9~8 Svhr, which is very low and close to background levels. When compared to the results of computer analysis, an error ranging from 27% to 97% was observed. It is concluded that validating the accuracy in the low dose rate range close to background levels is challenging through a comparison of calculated and measured dose rates.

저자
  • Jisoo Yoon(Korea Hydro and Nuclear Power (KHNP) Central Research Institute (KHNP-CRI)) Corresponding author
  • Kyung Rok Park(Korea Hydro and Nuclear Power (KHNP) Central Research Institute (KHNP-CRI))
  • Jung-Kwon Son(Korea Hydro and Nuclear Power (KHNP) Central Research Institute (KHNP-CRI))
  • Woo Yong Kim(KONES Corp.)