Appropriateness of the minimum detectable activity in the analysis of gamma radionuclides is very important. This is reason determine the time factor among the conditions of the analysis when it is rationally determined has the advantage that radioactivity analysis can be performed accurately and quickly. In this study, 100 mL of an unknown sample was diluted in Marinelli Beaker 1L to obtain, review data on gamma radiation analysis results and minimum detectable activity for each measurement time. The measurement was used High Purity Germanium detector, target nuclides are Co-57, Co-58, Y-88 and Cs-137. Since the radioactivity analysis sample will be expected to be the waste subject to selfdisposal or less during the radioactive waste classification, the minimum detectable activity standard was set based on the detection of less than the permissible activity for self-disposal for each nuclide. The measurement methods were measured by classifying it into seven categories: 1000 seconds, 3600 seconds, 10000 seconds, 30000 seconds, 80000 seconds, 100000 seconds, and 150000 seconds. The radioactivity from this measurement are Co-57 2.89 Bq·g−1, Co-58 0.19 Bq·g−1, Y-88 0.20 Bq·g−1, Cs-137 0.15 Bq·g−1, the measurement results under all conditions were similar. On the other hand, the minimum detectable activity showed values above the allowable activity for self-disposal in not but Co-58 at 1000 and 3600 seconds. Only after taking the measurement time of 10000 seconds, the result was derived Co-57 0.0095 Bq·g−1, Co-58 0.0068 Bq·g−1, Y-88 0.0052 Bq·g−1, Cs-137 0.0062 Bq·g−1, which was confirmed to less than the allowable activity for self-disposal by nuclide. Reasonably determining the measurement time in gamma radionuclide analysis is a very important issue in terms of economy of time and accuracy of measurement. Although this study cannot be said to be able to determine a reasonable measurement time for all gamma radionuclide analysis, it is hoped that research on various samples will be made to contribute to the efficient measurement of gamma radioactivity.
핵분열로 인해 생성되는 방사성 노블가스는 주변국의 핵활동을 감시할 수 있는 중요한 지표 핵종이다. 특히 제논은 생성량이 많고 반감기가 짧아 핵실험 탐지에 적합하며 크립톤은 핵연료 재처리 탐지의 추적자로 활용되고 있다. 방출된 방사성 노블가스는 막대한 대기에 희석되어 농도가 감소하고 일부는 시간에 따라 방사능이 감쇠하기 때문에 대기 중에는 매우 극미량으로 존재하게 된다. 따라서 측정을 통해 의미 있는 데이터를 얻기 위해서는 가능한 낮은 수준의 MDA를 설정하는 것이 중요하다. 본 연구에서는 방사성 제논과 크립톤을 동시 포집 할 수 있는 장비인 BfS-IAR시스템을 활용하여 이론을 통해 MDA 를 산출하였다. 또한 MDA 산출방식의 변화, 신뢰수준의 정도는 물론 계측 조건의 변화에 따른 영향을 확인하고 MDA를 저감하기 위한 방안들을 모색하였다. 그 결과 배경농도가 극미량인 제논의 경우 전처리과정의 효율화와 안정적인 계측 성능 유지가 가장 중요한 요소로 판단되었으며, 크립톤의 경우 제논과 달리 시료의 방사능이 높기 때문에 MDA 재설정을 통한 분석조건이나 시스템 최적화를 통해 효율적인 분석을 수행할 수 있을 것으로 판단된다.