In order to indirectly evaluate the inventory of difficult-to-measure (DTM) nuclides in radioactive waste, the scaling factor method by key nuclide has been used. It has been usually applied to low-and intermediate-level dry active waste (DAW), and the tolerance of 1,000% margin of error in the US, that is the factor of 10, is applied as an allowable confidence limits considering the inhomogeneity of the waste and the limitation of sample size. This is because the scaling factor method is based on economic efficiency. Confidence limits is the uncertainty (sampling error) according to predicting the mean value of the population by the mean value of the sample at 95% confidence level, reflecting the limitations of sample size (representation) with the standard deviation. If the standard deviation is large, the sample size can be increased to satisfy the allowable confidence limits. In the new nuclear power plants, the concentration of cesium nuclide (137Cs) in radioactive waste tends to be very low due to advances in nuclear fuel and reactor core management technology, which makes it very difficult to apply cesium as a key nuclide. In addition, it is inevitable to apply the mean activity concentration method, which reasonably and empirically derives the concentration of DTM nuclides regardless of key nuclide, when the correlation between key and DTM nuclides is not significant. The mean activity method is a methodology that applies the average concentration of a sample set to the entire population, and is similar to applying the average concentration ratio between key and DTM nuclides of a sample set to the population in the scaling factor method. Therefore, in this paper, the maximum acceptable uncertainty (confidence limits) at a reasonable level was studied when applying the mean activity concentration method by arithmetic mean unlike the scaling factor method which usually uses the geometric mean method. Several measures were proposed by applying mutatis mutandis the acceptable standard deviation in radiation measurement and the factor of 10 principle, etc., and the appropriateness was reviewed through case analysis.
This case study explained the korea expressway’s ex-post evaluation that focused on the direct and indirect effects of expressway, constructed for 40years. Expressway has been the foundation of South Korean economy and society since 1970. But there was a few research of the various effect of expressway. This case study tried to qualitatively analyze the direct and indirect effects of expressway. This case study suggested the expressway’s various effects separated by a 10year from 1970 to 2010 and calculated the monetary value of expressway network by transport network analysis. Also, it suggested the SA(Service area) of express network from 1970 to 2010 by ArcGis’s IDW methodology. Especially, the regional and industry development effect of expressway was calculated in this study by adapting the function of Cobb-Douglas production.
국제전기기술위원회의 문서 IEC 60601-1의 3판 규격과 IEC 60601-2-45의 개별 규격에서는 진단용 X선 장치에서 X선 피폭 선량 정보를 표시하고 그 정확성을 명시할 것을 권고하고 있다. 하지만 임상에서 사용 하는 부착형 공기커마 면적선량계는 교정에 어려움이 있다. 이에 본 연구에서는 임상에 적용 가능한 RQR 표준 선질을 이용하여 공기커마 면적선량계의 에너지 의존성과 정확도를 평가하였고 임상에서 간접 교정을 시행할 수 있도록 방법론을 마련하고자 하였다. RQR5의 표준 선질에서 시행한 시험에서 부착형 공기커마 면적선량계는 –7.5%의 오차를 나타냈고, RQR8의 표준 선질에서는 –10.3%의 오차를 나타냈으며 시험한 모든 RQR 선질에 대해 평균 절대오차는 8.30%±2.85%를 나타내 IEC 60580과 AAPM TG 190의 조건을 만족하였다. 본 연구에서 도출한 공기커마 면적선량계의 교정 방법은 임상에서 사용하는 공기커마 면적 선량계의 간접 교정법으로 사용할 수 있을 것으로 기대한다.
본 연구의 목적은 DR (digital radiography) 장비에서 팬텀을 이용하여 간접변환방식의 CsI:Tl 검출기와 Gd₂O₂S 검출기를 중심으로 두께가 두꺼운 흉부 팬텀과 중간 두께의 대퇴부 팬텀, 그리고 피사체의 두께가 얇은 손 팬텀의 영상을 획득하고 SNR과 CNR을 분석하여 검출기의 특성을 알아보고자 하였다.
피사체 두께 변화에 따른 SNR과 CNR을 측정한 결과 중간 두께의 대퇴부 팬텀과 두께가 얇은 손 팬텀을 촬영 하였을 때 SNR과 CNR은 Gd₂O₂S 검출기보다 CsI:Tl 검출기에서 높게 나타났음을 확인 할 수 있었다. 그러나 두께가 두꺼운 흉부 팬텀을 사용하였을 때는 Gd₂O₂S 검출기의 SNR이 80∼125 kVp 일 때와 CNR이 80∼110 kVp 일 때 CsI:Tl 검출기 보다 값이 높게 나타났고, 저관전압에서 고관전압으로 갈수록 SNR과 CNR은 모두 증가하였다. 중간 두께의 대퇴부 팬텀에서 CsI:Tl 검출기의 SNR과 CNR은 40∼50 kVp 일 때 증가하다 고관전압으로 갈수록 감소하는 것을 확인 할 수 있었고, Gd₂O₂S 검출기의 SNR과 CNR은 40∼60 kVp 일 때 증가하다 고관전압으로 갈수록 감소하는 것을 확인 할 수 있었다. 두께가 얇은 손 팬텀에서 CsI:Tl 검출기의 SNR과 CNR은 저관전압에서 감소하다가 고관전압으로 갈수록 증가하면서 100∼110 kVp에서는 감소하였고, Gd₂O₂S 검출기의 SNR과 CNR은 고관전압으로 갈수록 감소하는 것을 확인하였다. MTF는 0.5 ∼3 lp/mm에서 CsI:Tl 검출기가 Gd₂O₂S 검출기보다 6.02∼90.90%로 높음 보여주고 있고, DQE는 0.5∼3 lp/mm에서 CsI:Tl 검출기가 Gd₂O₂S 검출기보다 66.67∼233.33% 높음 보여주고 있다.
결론적으로 MTF와 DQE의 비교에서는 CsI:Tl 검출기가 Gd₂O₂S 검출기보다 높게 나타났지만, 두꺼운 흉부 팬텀에서 일정 관전압 구간에서는 저가의 Gd₂O₂S 검출기가 고가의 CsI:Tl 검출기보다 SNR과 CNR이 높다는 것을 확인하였다. 흉부 X선 검사 시 CsI:Tl 검출기보다 Gd₂O₂S 검출기를 사용하여 우수한 화질의 흉부 영상을 구현함으로써 검사에 유용할 것으로 판단되어지며, 사용자 입장에서 검출기 형태를 결정 할 때 가격 대비 성능을 고려 해볼 사항으로 판단된다.