The purpose of this study was to investigate the actual conditions of radiation safety supervision in animal clinics using inspection standard of X-ray generator for diagnosis. The surveys for inspection standard system, equipment condition, and safety supervision were carried out in 18 animal clinics randomly. The inspection standard included reproducibility of dose exposure, kVp, mAs, collimator accuracy test, collimator luminance test, X-ray view box luminance test, grounding system equipment test and external leakage current test. The surveys of equipment condition and safety supervision used one-on-one interview with 5 points measurement. As a result, 44.44% of reproducibility of dose exposure was proper, 81.25% of kVp test was good, and 100% of mAs test was appropriate. Also, 66.66% of collimator accuracy test was proper, 61.11% of collimator luminance test was good, 53.13% of X-ray view box luminance test was suitable. In addition, only 5.55% of grounding system equipment and ground resistance was proper, 63.64% of external leakage current test was appropriate in grounding system equipment test. The 100mA electric capacity of X-ray generator for diagnosis was popular with 44.44%, and its 55.56% was purchased used equipment. Monthly average of less than 50 times (61.11%) was top frequency in use, and no animal clinic had a thermo-luminescence dosimeter(TLD). The 16 animal clinics with radiation safety zone and 2 without radiation safety zone were appeared.
1895년에 독일 뢴트겐에 의해 발견된 이후 X선은 1896년부터 의료목적으로 사용된 기록이 있으며, 한국에 도입되어 진단용X선발생장치로서 환자 진료에 사용하기 시작한 역사는 1911년경부터 조선총독부의원과 세브란스의원에 도입되어 사용하게 되었다. 최근 의학 및 의용공학의 발전으로 방사선을 이용한 진단 기술의 발전과 건강에 대한 국민의 의료욕구가 증가함에 따라 방사선을 이용한 질병의 진단과 치료 방법에서 새로운 기법이 개발되고 그 이용은 지속적으로 증가되고 있는 추세이다. 방사선의 이용은 진단 및 치료에 중대한 이득을 제공하고 있으나 그 이면에는 최적화 및 정당화 되지 않은 방사선의 피폭으로 장애 및 유해요인이 초래되는 것은 부인할 수 없다. 이와 관련하여 방사선을 이용하여 검사 및 치료를 시행함에 있어 피폭선량 최적화를 위한 기술적 노력은 방사선 관계자의 막중한 의무이다.